1. 8.23 Liquid-in-glass thermometers are suitable for use in obtaining cargo product temperatures in marine vessels required to operate with closed or restricted systems? |
2. 8.36 A liquid-in-glass thermometer with a range between 60°F and 180°F and must be accurate to within? |
3. 8.42 In motion is defined as continuously raising and lowering the probe above and below the desired temperature measurement depth, by approximately how much? |
4. 8.21 To what increment should a portable electronic thermometer be read and recorded? |
5. 8.36 A liquid-in-glass thermometer with a range between 60°F and 180°F and must be accurate to within? |
6. 8.12 If a portable electronic thermometer has a range of 0 - 95 °C [32-200°F], what is the required resolution? |
7. 8.54 According to API MPMS Chapter 17.1, the preferred instrument for taking temperature is? |
8. 8.17 If a tank has more than 4.5 m [15 feet] of liquid, what is the minimum number of temperature readings that must be taken? |
9. 8.48 What is the recommended immersion time, for a liquid-in-glass thermometer in a cup-case assembly, in motion, in a product with density of 950 kg/m3 [API gravity 17.4] |
10. 8.09 What is the minimum acceptable accuracy for a portable electronic thermometer when it is indicating a temperature reading of 101 °C [215°F]? |
11. 8.11 Which standards address the earthing (grounding) of portable electronic thermometers? |
12. 8.32 When taking temperatures for ROB, OBQ or slops containing oil and water it is important ____ |
13. 8.45 What is the minimum number of temperatures to be taken in a tank containing less than 3m [10 feet] of product? |
14. 8.20 What is the minimum amount of product needed for a temperature to be taken? |
15. 8.38 How should scale graduation marks be applied to a liquid-in-glass thermometer? |
16. 8.47 When taking temperatures, in motion means to move the temperature measurement device approximately 60cm [2 feet] above and 60cm [2 feet] below the desired measurement location? |
17. 8.07 When verifying a portable electronic thermometer with a range of 0 to 95 °C [32 to 200°F] , what is the maximum permissible error before it should be recalibrated? |
18. 8.43 What does the term temperature stratification mean? |
19. 8.34 A thermowell used to measure temperature must be filled with a suitable heat-transfer liquid. |
20. 8.44 When using a liquid-in-glass thermometer in a cup-case assembly to measure a product with density of 904 kg/m3 [API gravity 24.9] what is the minimum length of time the thermometer should be immersed in the liquid when the assembly is in motion during the measurement process? |
21. 8.31 What is the maximum amount by which the reading on a portable electronic thermometer can vary when it may be considered to have stabilized? |
22. 8.39 An ASTM 59C [ASTM 59F-80] liquid-in-glass thermometer has a scale range of -18°C to 82°C [0°F-180°F]. What should the graduations on this thermometer be? |
23. 8.24 According to ISO 4268 [API Chapter 7] the minimum amount of time a stationary PET should stay in a product with a density greater than 935 kg/m3 is ? |
24. 8.01 API MPMS Chapter 7/ISO 4268 pertains to what type of thermometers? |
25. 8.05 How does calibration of a portable electronic thermometer differ from a field check? |
26. 8.37 Which API MPMS Chapter covers temperature determination? |
27. 8.16 What must be done immediately before taking temperatures using a portable electronic thermometer? |
28. 8.35 Which ISO standard covers temperature determination? |
29. 8.50 A 4 000m3 [25 000bbl ] capacity storage tank has a depth of product of 4.70 m [15 feet] in the tank. What is the minimum number of temperature measurements that should be obtained on this tank. |
30. 8.25 According to API MPMS Chapter 7 the minimum amount of time that an in motion probe should stay in a product with an API gravity less than 20 is? |
31. 8.26 If a tank has an innage of 2,8m [9 11 ] what is the minimum number of temperatures which should be taken? |
32. 8.49 What temperature is to be used during an ROB survey if there is only 5 cm [2 inches] of liquid in the bottom of the tank? |
33. 8.51 The immersion time for a cup-case assembly is minimized by continually raising and lowering the assembly 0.3m [1 foot] above and below the desired temperature measurement point. |
34. 8.33 When using a portable electronic thermometer, what is the minimum amount of time the probe must stay in a middle distillate of density 821.7 kg/m3 [API gravity 40.7] , if the probe is moving? |
35. 8.10 To what increment should the display of a portable electronic thermometer be capable of reading? |
36. 8.53 What is the purpose of keeping the probe of a portable electronic thermometer (PET) in motion? |
37. 8.27 On a portable electronic thermometer, at what temperature should the field check be performed ? |
38. 8.03 Is a portable electronic thermometer required to have a low voltage indicator? |
39. 8.13 Upright cylindrical storage tanks have calibration tables based on a specific tank shell temperature. If the observed tank shell temperature differs from the calibration table tank temperature will the volumes extracted from the calibration table will need to be corrected for this temperature difference? |
40. 8.18 If only one temperature is required where should this temperature be taken from? |
41. 8.02 Does API MPMS Chapter 7/ISO 4268 make any references to the construction of portable electronic thermometers? |
42. 8.06 On a portable electronic thermometer, what should be checked at least once per month? |
43. 8.04 Why do portable electronic thermometers have low-voltage indicators? |
44. 8.46 When taking the temperature of a product in a tank and there is more than 1°C [2 °F] difference between upper, middle and lower temperatures, what steps should be taken? |
45. 8.52 What procedure is used to field check a liquid-inglass thermometer in a cupcase assembly? |
46. 8.40 When first received from the manufacturer or equipment supplier, what should happen to a new liquid-in-glass tank thermometer? |
47. 8.28 When using a portable electronic thermometer what is an indication of temperature stabilization? |
48. 8.15 What information must be recorded when taking a temperature of a cargo using an in-line probe ? |
49. 8.41 A liquid-in-glass thermometer must be verified against a thermometer traceable to a national standard when new and then at maximum intervals of? |
50. 8.08 To what precision should temperatures obtained using portable electronic thermometers be read and recorded? |
ให้ตรวจสอบชื่อนามสกุลว่ากรอกครบหรือไหม่
พร้อมกับตรวจข้อสอบว่าเลือกทำครบทุกข้อหรือไม่ ไม่ครบไม่ตรวจนะครับ
|
|
|